Pipette Calibration Calculator
Verify pipette accuracy and precision using the gravimetric method (ISO 8655) with Z-factor temperature correction.
Inputs
Results
Overall pass (1=yes, 0=no)
0
Mean dispensed volume (μL)100.27
Accuracy error (μL)0.27
Accuracy error (%)0.27
CV% (precision)0.21
Standard deviation (μL)0.21
Range (μL)0.5
Z-factor (μL/mg)1
ISO accuracy limit (%)0.8
ISO precision limit (%)0.2
Accuracy pass (1/0)1
Precision pass (1/0)0
Z Factor1
How to Use This Calculator
- Enter the pipette's nominal volume setting in microliters.
- Input the measured gravimetric weight in milligrams from 10 replicate dispenses.
- Set the water temperature for density correction (typically 20-25°C).
- Review the calculated mean volume, systematic error (%), and CV% for precision.
- If error exceeds ±2% or CV exceeds 1%, send the pipette for recalibration before use.